5
Subassembly Removal and Replacement
Chamber Module
Electrode Assembly (03-51157) Spacers and Slides (35-51152),
Lower Shelf Stop (35-51153), Upper Shelf Stop (33-51158)
Screw
Screw
Figure 27. Electrode Assembly
Required Tools:
2.5 mm hex wrench.
Preparation:
1. Turn the sterilizer off.
2. Remove the shelves from the chamber.
Removal:
4
Note
: You may find it easier to remove the entire
electrode when you are replacing the plastic items.
1. Remove 2 screws attaching the electrode to the front of the
chamber.
2. Remove the screw and lock washer near the bottom center of the
electrode connecting the electrode to the feedthrough.
3. Gently pull the electrode assembly straight out of the chamber to
remove.
4. Remove the spacers and shelf stops.
Replacement:
Replacement is the reverse of the removal steps.
84 STERRAD
®
NX
™
Service Guide