Maintenance
6
2. Install a new optical window and O-ring in the lamp housing (if
necessary).
3. Clean the inside of the Ultratorr fitting. Lightly coat a new O-ring
with approved vacuum grease and place the O-ring in the Ultratorr
fitting. Reinstall the spacer in the Ultratorr fitting.
4. Place the threaded collar around the base of the lamp housing. Insert
the lamp housing tube into the Ultratorr fitting. Press down until it is
seated firmly.
5. Orient the lamp housing so that there is no interference with the lamp
cable. Use soft-jaw pliers to tighten the collar.
6. Reconnect cable P20.
Delivery Subsystem Inlet Filter
Replacement
1. Slide the carriage on its lead screw toward the rear of the sterilizer to
cover the piercing needles.
2. Unscrew and disconnect the inlet filter cartridge.
To install the new inlet filter, do the following:
1. Connect a new inlet filter to the inlet fitting.
2. Move the carriage back into position
Air Pump Tubes Replacement
1. Unscrew the air pump from the bracket.
2. Remove the two tubes on the air pump.
To install the new tubes, do the following:
1. Place the tubes on the outlet and inlet ports.
2. Replace the air pump on the bracket making sure it is securely
fastening.
STERRAD
®
NX
™
Service Guide
105