5
Subassembly Removal and Replacement
Power Inlet Connector (45-51407)
Screw
Screw
Insulating
Cover
Inlet
Connector
Retainer
Figure 25. Power Inlet Connector
WARNING! ELECTRIC SHOCK HAZARD. HAZARDOUS
VOLTAGES ARE PRESENT AT THE POWER INLET
CONNECTOR. TURN THE STERILIZER OFF AND UNPLUG
IT FROM THE MAINS BEFORE SERVICING THE POWER
INLET CONNECTOR.
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Unplug the sterilizer from the main power source.
3. Remove the front top panel, the top panel, and the rear top panel.
Removal:
1. Pull off 3 push-on type lead connections to the rear of the power
inlet. Note the color-coded location of each lead.
2. Remove the two screws that attach the power inlet to the chassis.
3. Pull the power inlet away from the rear panel to remove.
Replacement:
Replacement is the reverse of the removal steps.