Subassembly Removal and Replacement
5
Door Assembly (03-51985)
The alignment of the door (and the latch) must be performed by adjusting
screws located in the top and bottom door hinges. The door must be
removed to gain access to these two adjustment screws.
Grounding
Wire Screw
Hinge
Hinge
Figure 28. Door Assembly
WARNING! HOT SURFACES. THE INNER SURFACE OF
THE DOOR IS HOT (50º C) WHEN ENERGIZED. DO NOT
TOUCH THE DOOR WHEN THE SYSTEM IS OPERATING.
TURN OFF THE STERILIZER AND ALLOW IT TO COOL
BEFORE PERFORMING SERVICE ON THE DOOR.
Required Tools:
2.5 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, front bottom panel, top panel, and
left side panel and door cover.
Removal:
1. Remove the screw holding the cable strap.
2. Unplug cables: P2 and P138.
3. Remove the screw holding the grounding wire lug to the door.
4. Pull the cables out of the door through the cutout on the left end
of the door.
5. Hold the top of the door. Release the catch holding the top door
hinge to the hinge bracket by pressing the catch lever inward
toward the chamber.
STERRAD
®
NX
™
Service Guide
85