5
Subassembly Removal and Replacement
6. Hold the door carefully. Release the catch holding the bottom
door hinge to the hinge bracket by pressing the catch lever inward
toward the chamber.
7. Pull the door straight out from the front of the chamber until the
hinges are free of the bracket.
Replacement:
Replacement is the reverse of the removal steps. The following
alignment procedure is necessary to make sure the door functions
correctly.
Door Alignment
Alignment
Using a right-angle ratchet Phillips screwdriver, adjust the door position
by turning adjustment screws on the hinge. The door is properly adjusted
when the following criteria are met:
•
The latch on the right side of the door is centered left-to-right and
top-to-bottom with the opening in the catch mechanism. The latch
should enter the catch without making contact with the sides of the
catch opening.
•
When the door is closing, the left side of the door should first make
contact with the chamber O-ring at the same time as the latch begins
to enter the catch mechanism.
•
Adjust the front-to-back tilt of the door so that the door contacts the
top and bottom of the chamber O-ring at the same time. The door
should not make contact with the bottom portion of the O-ring while
there is a gap between the door and the top portion of the O-ring, nor
should it make contact with the top portion of the O-ring while there
is a gap at the bottom.
•
With the door closed, check around the entire perimeter of the door
to make sure there are no gaps between the door and the O-ring.
86 STERRAD
®
NX
™
Service Guide
Содержание sterrad NX
Страница 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Страница 10: ......
Страница 16: ......
Страница 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Страница 38: ......
Страница 100: ......
Страница 124: ......