![Agilent Technologies 4339B Скачать руководство пользователя страница 215](http://html.mh-extra.com/html/agilent-technologies/4339b/4339b_operation-manual_2867889215.webp)
4339B
Resistivity
Resistivity
V
olume
Resistivity
In
general,
volume
resistivity
is
dened
as
the
resistance
per
unit
volume
.
v
=
Ar ea
t
R
v
2
1
10
where
,
v
V
olume
resistivity
[-cm]
Area
Eective
area
[mm
2
]
t
Thickness
of
the
sample
[mm]
R
v
Measured
volume
resistance
[]
When
the
16008B
Resistivity
Cell
is
used,
the
above
equation
becomes:
v
=
2
D
1
+
B ( D
2
0D
1
)
2
2
4t
2
R
v
2
1
10
Where
,
D
1
Main
electrode
diameter
[mm]
D
2
Guard
electrode
diameter
[mm]
B
Eective
area
coecient
B
is
the
fraction
of
the
gap
width
to
be
added
to
the
diameter
of
the
circular
electrodes
or
to
the
dimensions
of
rectangular
or
cylindrical
electrodes
.
The
value
is
usually
specied
by
the
standard
that
you
use
.
(F
or
example:
1
for
ASTM
D
257;
0
for
JIS
K6911).
Figure
7-1.
V
olume
Resistivity
Measurement
Basics
7-3
Содержание 4339B
Страница 10: ......
Страница 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 30: ......
Страница 44: ......
Страница 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Страница 212: ......
Страница 220: ......
Страница 230: ......
Страница 256: ......
Страница 262: ...4339B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 263: ...4339B Figure B 2 Handler Interface Control Output Signals Diagram Handler Interface Installation B 3 ...
Страница 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...