Agilent 4339B High Resistance Meter Operation Manual
Manual Change
Agilent Part No. N/A
Sep 2008
Change 1
Change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to the following.
Source
Voltage
Setting
Test Limits
Test Result
1
Measurement
Uncertainty
0 V
±0.10 V
V
± 0.000206 V
10 V
±0.12 V
V
± 0.00027 V
25 V
±0.14 V
V
± 0.00054V
50 V
±0.18 V
V
± 0.0010 V
100 V
±0.26 V
V
± 0.0019 V
200 V
±0.42 V
V
± 0.0037 V
201 V
±0.82 V
V
± 0.0040 V
250 V
±0.90 V
V
± 0.0053 V
500 V
±1.3 V
V
± 0.0097 V
1000 V
±2.1 V
V
± 0.0188 V
Change 2
Change the Measurement Uncertainty value of Ammeter Offset Voltage Test (Page 9-19) to the following.
Test Limits
Test Result
1
Measurement
Uncertainty
±0.5 mV
mV
± 0.0083 mV
Change 3
Change the Measurement Uncertainty value of Ammeter Input Resistance Test (Page 9-19) to the
following.
Test Limits
Test Result
1
Measurement
Uncertainty
±50
ohm
ohm
± 4.37
ohm
Содержание 4339B
Страница 10: ......
Страница 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 30: ......
Страница 44: ......
Страница 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Страница 212: ......
Страница 220: ......
Страница 230: ......
Страница 256: ......
Страница 262: ...4339B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 263: ...4339B Figure B 2 Handler Interface Control Output Signals Diagram Handler Interface Installation B 3 ...
Страница 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...