4339B
CALibration
Subsystem
COMMAND
P
ARAMETER
CALibration
[:ALL]?
:AUTO
<Boolean>
:CALibration[:ALL]?
P
erforms
a
calibration,
and
then
returns
an
error
code
.
The
error
code
is
referred
to
\Messages"
at
the
back
of
this
manual
(An
error
code
of
\0"
indicates
no
failure).
:CALibration:A
UTO
fON j OFF j 1j 0g
Sets
or
queries
if
the
oset-error
canceling
function
is
enabled.
The
oset-error
calibration
function
cancels
the
oset
error
due
to
the
temperature
change
,
when
the
measurement
range
is
100
pA
or
1
nA
and
the
measurement
time
mode
is
Long.
ON ,
1
enables
oset-error
canceling
OFF ,
0
disables
oset-error
canceling
Query
response
is
1
or
0
5-16
GPIB
Reference
Содержание 4339B
Страница 10: ......
Страница 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 30: ......
Страница 44: ......
Страница 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Страница 212: ......
Страница 220: ......
Страница 230: ......
Страница 256: ......
Страница 262: ...4339B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 263: ...4339B Figure B 2 Handler Interface Control Output Signals Diagram Handler Interface Installation B 3 ...
Страница 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...