![Agilent Technologies 4339B Скачать руководство пользователя страница 161](http://html.mh-extra.com/html/agilent-technologies/4339b/4339b_operation-manual_2867889161.webp)
4339B
CALCulate
Subsystem
:CALCulate1:MA
TH:EXPRession:CA
T
alog?
Returns
available
parameters
which
are
used
with
CALCulate1:MATH:EXPRession:NAME
command.
(query
only)
F
or
more
information,
see
CALCulate1:MATH:EXPRession:NAME
command.
Query
response
is
{DEV|PCNT} .
:CALCulate1:MA
TH:EXPRession:NAME
f
DEV
j
PCNT
g
Denes
or
queries
the
expression
used
for
deviation
measurement,
which
is
enabled
by
CALCulate1:MATH:STATe.
The
reference
value
can
be
dened
using
the
DATA[:DATA]
command.
DEV
Absolute
value
of
deviation
PCNT
P
ercentage
of
deviation
:CALCulate1:MA
TH:ST
A
T
e
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
if
math
processing
dened
by
:CALCulate1:MATH:EXPRession:NAME
is
enabled
or
not.
ON ,
1
Enables
math
processing.
OFF ,
0
Disables
math
processing.
Query
response
is
1
or
0 .
:CALCulate1:P
A
TH?
Returns
the
order
in
which
the
CALCulate
subsystem
commands
are
to
be
performed.
(Query
only)
The
4339B
always
processes
measured
data
in
order
of
:CALCulate1:FORMat
subsystem,
:CALCulate1:MATH
subsystem,
and
:CALCulate1:LIMit
subsystem.
Query
response
is
always
FORM,MATH,LIM
.
:CALCulate1:RESistivity:EARea
<numeric
value
>
Species
or
queries
the
eective
area
of
the
main
electrode
for
the
resistivity
measurement.
<numeric
value>
is
,
numeric
0
to
0.99999
(m
2
)
Query
response
is
a
numeric
value
in
<NR3>
format.
:CALCulate1:RESistivity:EPERimeter
<numeric
value
>
Species
or
queries
the
eective
perimeter
of
the
electrode
for
the
resistivity
measurement.
<numeric
value>
is
,
numeric
0
to
9.9999
(m)
Query
response
is
a
numeric
value
in
<NR3>
format.
GPIB
Reference
5-13
Содержание 4339B
Страница 10: ......
Страница 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 30: ......
Страница 44: ......
Страница 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Страница 212: ......
Страница 220: ......
Страница 230: ......
Страница 256: ......
Страница 262: ...4339B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 263: ...4339B Figure B 2 Handler Interface Control Output Signals Diagram Handler Interface Installation B 3 ...
Страница 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...