![Agilent Technologies 4339B Скачать руководство пользователя страница 211](http://html.mh-extra.com/html/agilent-technologies/4339b/4339b_operation-manual_2867889211.webp)
4339B
Measuring
Insulation
Resistance
Time
Characteristics
9.
Set
sequence
mode
parameters
.
Press
.
a.
Set
the
charge
time
.
Select
Chrg .
Enter
the
charge
time
value
,
in
this
example
60
(s).
Press
.
b.
Set
the
interval
time
and
memory
size
.
Select
Intvl .
Enter
the
interval
time
value
,
in
this
example
10
(s).
Press
.
Then
enter
the
number
of
measurements
(equivalent
to
the
memory
size),
in
this
example
60.
Press
.
c.
Select
the
continuous
mode
as
the
measurement
sequence
mode
.
Press
.
Select
Continuous
using
or
and
press
.
10.
Set
the
4339B
to
T
alk
Only
mode
.
Press
and
enter
31 .
Press
.
11.
Start
the
measurement
sequence
.
Press
.
The
printer
begins
printing
data.
d
a
+0,+6.10,+0
+0,+6.10,+0
+0,+6.10,+0
.
.
.
Figure
6-7.
Printed
Results
Note
If
you
want
to
abort
the
measurement
sequence
,
press
.
12.
When
the
measurement
sequence
is
completed,
change
the
GPIB
address
to
an
address
other
than
31.
13.
W
ait
until
the
High
V
oltage
indicator
is
OFF
.
Then
remove
the
DUT
from
the
test
xture
.
Application
Measurement
6-11
Содержание 4339B
Страница 10: ......
Страница 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 30: ......
Страница 44: ......
Страница 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Страница 212: ......
Страница 220: ......
Страница 230: ......
Страница 256: ......
Страница 262: ...4339B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 263: ...4339B Figure B 2 Handler Interface Control Output Signals Diagram Handler Interface Installation B 3 ...
Страница 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...