2D-LC User Guide
89
4
Installation
Software Installation
Software Installation
2D-LC Software Installation in Agilent Masshunter
Workstation
Prerequisites
A compatible CDS must be installed first. For details, see the CDS
documentation.
To observe if your computer full fills the requirements, like for the hardware CPU,
memory, hard disk space and the software, check the windows settings.
It is recommended to use the Windows 10 Professional for MassHunter
Workstation.pdf (p/n G3336-90036) as guidance for the installation.
•
Set up the computer system
•
Check PC network card configuration
•
Prepare for installation
•
To make sure you have the latest critical updates and security fixes, run
Windows Update
•
Make sure that Windows Update is completed before you continue
1
Install the Data Acquisition program.
2
Install the Qualitative Analysis program.
3
Install the Quantitative Analysis program.
[OPTIONAL]
4
Install Microsoft Excel.
NOTE
•
If you are upgrading from a Q-TOF 10.x build, uninstall 10.x first.
•
Be sure that there has been no other MassHunter installations on the PC or
you will need to re-image.
•
Be sure
ALL Windows Updates
are COMPLETED before Installing a build the
first time or you may have to re-image again.
•
Decide on the installation type.
This decision is permanent
.
If you change your mind later, you will have to re-image.
NOTE
For 2D-LC setups only the noncompliant workstation that includes 2D-LC
support will work. Workstation Plus and Enterprise will not support 2D-LC.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...