2D-LC User Guide
213
8
Run the System
Run the Experiment
1
Load method
High Resolution Checkout
from the 2D-LC data media and
modify the settings for your multiple heart cutting configuration.
2
Run the method with 2D-LC starter sample, 1 x 2 mL (5190-6895), 1:10 (for
only UV Checkout), 1:100 (for LCMS Checkout), or 1:1000 (for LCMS
Checkout) diluted with Methanol/Water (20/80; v/v) with 0.1 % formic acid.
3
If necessary, subsequently edit or optimize the method.
Min Height
1000 counts
Reference Masses
Positive
121.05087300
922.00979800
Chromatograms
Chrom Type Label Offset Y-Range
TIC TIC 1510000000
TIC TIC 1510000000
Stop time
As pump/No limit
Table 23
Recommended conditions in 2D (LC-MS)
Parameter
Value