![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 372](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575372.webp)
14
Theoretical Background
2D as detector
2D-LC User Guide
372
Figure 175 Principle of 2D-LC (example for LCxLC): Effluent of first column (1) is sampled (2) and
injected to second column (3). Peaks of second column separation are detected and
reconstructed.
1st peak from
1st dimension
3rd peak from
1st dimension
First dimension
Second dimension
(1)
(2)
(3)
(1)
(3)
Peak reconstruction
(software based)
V
isualization
2nd peak from
1st dimension
(1)
First dimension separation
(2)
Sampling of the first dimension
(3)
Second dimension separation
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...