4
Installation
Hardware Installation
2D-LC User Guide
84
Figure 40
Setup G. Single Heart-Cutting Configuration as Single Sample Loop Setup
G
Number of
Capillary
#
Connection
ID x L [mm]
P/N
Description
1
1
Bypass capillary (OUT)
0.12 x 105
5500-1238
Capillary, ST 0.12x105 SL/SL
2
1
Waste line
self-cut x 0.7
0890-1713
Tubing-flexible 0.8/1.61mm PTFE
WT (delivered with UV detector)
3
1
From pressure release kit; from
1
D column,
1
D detector
0.17 x 105
0.12 x 500
5500-1240
5500-1157
Capillary ST 0.17x105 SL/SL
Capillary ST 0.12x500 SL/S
4
Sample Loop (IN)
5004-0036
180 µL Loop 2D-LC as an example
5
Sample Loop (OUT)
5004-0036
180 µL Loop 2D-LC as an example
6
1
To
2
D column
0.12 x 400
5500-1251
Capillary ST 0.12x400 SL/SL
7
1
From
2
D pump
0.17 x 280
5067-4608
Capillary ST 0.17x280 SX/S
8
Bypass capillary (IN)
0.12 x 105
5500-1238
Capillary, ST 0.12x105 SL/SL
For all other capillaries / connections, see
on page 74, and
NOTE
If the dual-loop setup has been selected in the software configuration (see
install mirror-inverted, the sample loop at port 1 and 8 and the bypass capillary at position 4 and 5.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...