Concepts of 2D-LC
Heart-Cutting 2D-LC (LC-LC)
2D-LC User Guide
18
Heart-Cutting 2D-LC (LC-LC)
The following items are characteristic for LC-LC:
•
Only parts of the effluent of the
1
D column - only the peaks of interest eluted
from the
1
D column - are injected to the
2
D column
•
A peak from the first dimension is sampled as a whole and a method with a
lower flow rate and a gradient typically with a longer run time than the
collection time is used to improve separation efficiency
•
Typically longer columns with higher separation efficiency are used in
2
D
column
NOTE
Heart-Cutting 2D-LC (LC-LC) is the method of choice if the samples to analyze
are known or to improve confidence of an experiment (pharma, method
development and so on).
NOTE
Multiple peaks eluted from the first dimension column can be sampled and
analyzed in the second dimension but the run time of the second dimension
must match the retention time between two first dimension peaks.
A started
second dimension analysis will always be finished!
Thus, a second peak being
eluted from the first dimension might be lost, if sampled while the second
dimension analysis is still running.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...