4
Installation
Hardware Installation
2D-LC User Guide
64
Single Heart-Cutting Configuration
2D-LC instruments that are exclusively used for Single Heart-Cutting and
Comprehensive 2D-LC experiments only require the standard 2D-LC valve
(G4236A). The valve can be conveniently attached to any Infinity II pump that is
installed. For a SHC configuration, transfer capillaries (6a/6b) are not necessary
since MHC decks are not installed.
Figure 26
Schematics of a Single Heart-Cutting (SHC) Configuration with supported valves. For
technical reasons, the ASM valve (G4243A) is not supported in Single Heart-Cutting
setups.