2D-LC User Guide
221
8
Run the System
Run the Experiment
Sampling Table
Start 0.2 min Time based, 2.5 min Peak Based, 4.50 min
2D Gradient
Analysis 1.50 min Equilibration 0.50 min Cycle time 2.12 with ASM ON and ASM Factor 3
Analytical gradient - Shifted Gradient Shift 1D:
3 % B 0.00 min
3 % B 0.37 min
10 % B 0.38 min
60 % B 1.62
30 % B 6 min
2
D Detector (DAD)
Diode-array
254 nm, Bandwidth 4 nm
Reference Wavelength 360 nm
Reference Bandwidth
100 nm
Peak width
80 Hz
Stop time
As pump/No limit
Table 29
Recommended conditions in 2D (HPLC) for ASM MHC 2D-LC
Parameter
Value