Typically very complex samples are analyzed by comprehensive 2-dimensional
liquid chromatography. The compounds which are often co-eluting form the first
dimension are further separated in the second dimension. With the Agilent 1290
Infinity II 2D-LC Solution always one large data-file spanning the run-time of the
two-dimensional analysis will be acquired. As an example, a 2-dimensional
analysis of a mixture of 26 polyphenolic standard compounds is shown in a one
dimensional data analysis display (
on page 257). Theoretically, the
data can be analyzed with Agilent MassHunter Workstation Software Qualitative
Analysis software.
But for easier data-analysis and better visualization of the comprehensive 2D-LC
data special software is recommended. Agilent recommends GC Image LCxLC
edition Software from GC Image LLC, Nebraska, USA. A trial download can be
found on
www.GCImage.com
as well as an online manual. Agilent 2D-LC data
files also including UV spectra and mass spectra data can be directly imported.
This software, with the information of the modulation time, is capable to extract
the data and isolate each second dimension run. Data will be reconstructed in a
two-dimensional display of the retention times. This can be displayed as a
colored 2-dimensional map of compound peaks (
baseline correction the peaks can be automatically detected by a peak detection
algorithm inherent in the 2D-LC data analysis software (
Since the third dimension is the intensity of the peaks a 3-dimensional plot of the
data is possible (
on page 258). With the given data set further
qualitative and quantitative data analysis is possible.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...