Installation
Hardware Installation
2D-LC User Guide
60
Recommendations for Instrument Setup
General Information
InfinityLab 2D-LC Solutions come in several flavors, still allowing flexible HPLC
combination of InfinityLab Series and 1200 Series Infinity modules. In
combination with the Agilent Mass Spectrometer the HPLC part of the 2D-LC
solution requires a two-stack configuration. For 2D-LC, a two-stack configuration
is always preferred. On the left stack, the order of the modules from bottom to
top is: pumps for both dimensions, then Vial- or Multisampler.
The sampler must be placed on top of the pumps. The right stack consists of one
or two column compartments and one or two standard UV detectors.
Depending on the number of solvents used, both stacks offer the possibility to
place a solvent cabinet on top.
Figure 24
Left: Recommended stack configuration for the 1290 Infinity II 2D-LC System. Right:
Bench space requirements of the 1290 Infinity II 2D-LC System.
NOTE
The dual stack configuration for 2D-LC requires at least 97 x 62 cm (24.4 x
38.2 inches) free, vertical bench space. 2.5 cm (1.0 inches) of space on either
side and approximately 8 cm (3.1 inches) in the rear is reserved for air circulation
and electric connections.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...