![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 365](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575365.webp)
2D-LC User Guide
365
14
Theoretical Background
Theoretical basis of 2D-LC
Resolution
A chromatographic separation can be optimized based on physical parameters
of the HPLC column such as particle size, pore size, morphology of the particles,
the length and diameter of the column, the solvent velocity, and the temperature.
In addition, the thermodynamics of a separation can be considered and the
properties of the solute and the stationary and mobile phases (percentage of
organic solvent, ion strength, and pH) can be manipulated to achieve the shortest
possible retention and highest selectivity.
1D-LC
Resolution (R
S
) can be described as a function of three parameters:
•
Column efficiency or theoretical plates (N),
•
Selectivity (
),
•
Retention factor (k).
Figure 170 Resolution equation
This means that the selection of appropriate mobile and stationary phase
properties and temperature is critical in achieving a successful separation.
Resolution in a one-dimensional separation usually is measured with:
Following results of this formula are important in practice:
•
R > 1.5
Peaks are completely baseline resolved
•
R > 1
Difference in retention time is larger than peak broadening, and therefore peak
spacing is adequate to observe distinct component zones
•
R < 0.5
Peaks are completely fused
R = Resolution
Δ
t = Difference in retention time maxima of two components
= Average standard deviation of two Gaussian peaks
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...