Error Information
General Error Messages
2D-LC User Guide
308
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Probable cause
Suggested actions
1
Leak detected in another module with a CAN
connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down
condition.
4
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error
condition. Refer to the
Service Manual
for the
degasser or the pump that has the degasser
built-in.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...