![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 324](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575324.webp)
12
Maintenance
Replace Valve Heads
2D-LC User Guide
324
Replace Valve Heads
Replace Valve Heads (G7116B)
Several optional valve heads are available, which can be installed and exchanged
easily.
Parts required
Description
Agilent Quick Change Valve Head.
CAUTION
The valve actuator contains sensitive optical parts, which need to be protected
from dust and other pollution. Pollution of these parts can impair the accurate
selection of valve ports and therefore bias measurement results.
Always install a valve head for operation and storage. For protecting the
actuator, a dummy valve head (part of Transportation Lock Kit
(G1316-67001 )) can be used instead of a functional valve. Do not touch
parts inside the actuator.
CAUTION
Column Damage or Bias Measurement Results
Switching the valve to a wrong position can damage the column or bias
measurement results.
Fit the lobe to the groove to make sure the valve is switched to the correct
position.
CAUTION
Valve Damage
Using a low pressure valve on the high pressure side can damage the valve.
When using multiple column compartments as part of a method
development solution, make sure that the high pressure valve head is
connected to the autosampler and the low pressure valve head is connected
to the detector.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...