10
Troubleshooting and Diagnostics
Lab Advisor Service
&
Diagnostic
2D-LC User Guide
288
Pump Leak Rate Test for the
2
D Pump
The test determines the leak rates in the primary and the secondary pump
chambers for component level diagnostic.
This 2D-LC test works only for the driver-based 2D-LC solution.
1
Select
Service & Diagnostic
from the menu.
2
Select the
2
D pump.
3
Select the
Pump Leak Rate Test
.
Figure 151 Pump Leak Rate Test for the 2D-LC 1290 binary pump
4
Press the
Run
button and follow the instructions in the software.
When
Diagnostic of the
2
D pump.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
•
Documentation provided with the Agilent Lab Advisor online help
•
2D-LC Manual
Procedure to follow:
•
Close the current Acquisition client window
•
Close instrument connection from OpenLab Control Panel
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...