Concepts of 2D-LC
Triggering of 2D-LC
The valve switches under the following conditions (whichever comes first):
•
If the
Sampling time
has elapsed, or
Figure 11
Peak triggering concept (elapsed sampling time)
•
If the signal falls below threshold or slope.
Figure 12
Peak triggering concept (signal falls below threshold or slope)