Troubleshooting and Diagnostics
Lab Advisor Service
&
Diagnostic
2D-LC User Guide
286
Lab Advisor Service
&
Diagnostic
Decluster the 2D-LC Cluster
This tool allows to remove an LC device's clustering configuration data, e.g. the
linking between
2
D pump and 2D-LC valve.
1
Select
Service & Diagnostic
from the menu.
2
Select the
2
D pump.
3
Select
Firmware Declustering
.
4
To
Clear clustering configuration data
, press the
Run
button.
Figure 149 Firmware Declustering of the 2D-LC 1290 binary pump
When
Replacement of one of the cluster partners.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
•
Documentation provided with the Agilent Lab Advisor online help
•
2D-LC Manual
Procedure to follow:
•
Close the current Acquisition client window
•
Close instrument connection from OpenLab Control Panel
NOTE
To re-establish the link between the two modules, re-perform an
Auto
configuration
and a selection as cluster.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...