2D-LC User Guide
27
2
Concepts of 2D-LC
Multiple Heart-Cutting and High Resolution Sampling 2D-LC
•
Cut number 5 cannot be parked entirely in the sample loop, otherwise cut 6
would go partially to the transfer capillary and would therefore be lost or spoil
cut 5
•
Cut 5 stays partially in transfer line and is immediately analyzed in
2
D