![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575088.webp)
4
Installation
Hardware Installation
2D-LC User Guide
88
Activate the 2D-LC System Driver With a Dongle-Based License
When you purchase Agilent 1290 Infinity II 2D-LC Solution from Agilent you will
receive a single USB stick which includes the 2D-LC dongle license. To run the
system and use its functionality, the
2
D pump must be activated. For this
purpose, the physical device is connected to the USB-port on the back of the
2D-LC pump. This will activate and enable the software to use the Agilent 1290
Infinity II 2D-LC Solution in your CDS.
Activate Agilent 1290 Infinity II 2D-LC Solution
1
Power off the module.
2
Plug the USB Dongle into the 1290 Infinity I or II binary pump USB Socket on
the back of the module.
3
Power on the module.
4
Once restarted, the 2D-LC License is activated and you can remove the USB
Dongle and store in a safe place.
Deactivate the License (Deactivation Steps in LabAdvisor)
For the deactivation of the license on the 2D-LC pump (e.g. you want to use the
license on a different 2D-LC pump) you have to use the LabAdvisor Diagnostic
Software.
1
Insert the USB Dongle at the rear of the
2
D pump.
2
Deactivate the license under
Instrument Control > Pump > Special Commands >
License Dongles
.
3
Remove the USB dongle at the rear of the pump and keep it on a safe place.
For further information, see
“Agilent Lab Advisor Software”
Parts required
Description
USB Dongle
NOTE
The dongle is required for a re-activation of 2D-LC License after mainboard
replacement.
NOTE
When the 2D-LC Driver connects to the instrument, it checks if a license is
available. If no licence is available, the driver remains offline. Tooltip when
hovering over the 2D-LC UI in the dashboard of the CDS shows the text:
No 2D-LC
license available
.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...