Troubleshooting and Diagnostics
Recommended Tests to Conclude Troubleshooting
2D-LC User Guide
304
Recommended Tests to Conclude
Troubleshooting
The following table shows the most important tests to conclude troubleshooting.
•
For further detailed information, see:
•
Maintenance information in the specific manual of each module.
•
Troubleshooting Guide
poster 5994-0709EN.
•
Best Practice for Using an Agilent LC System
01200-90090.
•
For additional help, contact your local Agilent Technologies service
representative.
Table 36
Recommended Tests for 2D-LC System Troubleshooting
Pump
Column
Compartment
Autosampler
Valve
Detector
2D-LC Instrument
Pressure Test
Leak Test
Thermostat Test
Pressure Test (if
column valve is
present)
Pressure Test
Inject standards or
inject different
volumina or blanks
Switching valve
position/Check
pressure reading
Pressure Test
Lamp Intentity
Test
Wavelength
calibration
In addition there
are detector
specific tests.
Run Checkout
For 2D-LC
Instruments
•
Pressure test
of the 1D-LC
Part
•
Pressure Test
of the 2D-LC
Part
Pump characteristic
•
Pump Ripple (1260 Pump)
•
Tuning (1290 pump)
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...