2D-LC User Guide
227
8
Run the System
Run the Experiment
Table 34
Recommended conditions in 2D (LC-MS)
Parameter
Value
Ion Source
Atmospheric pressure electrospray (Dual AJS ESI)
Ion Mode
Dual AJS ESI
Ion polarity
Both
Storage Mode
Both
LCMS Stream
MS
Acquisition Mode
Acquisition Mode MS1 Min Range (m/z) 50, Max Range (m/z) 1000, Scan
Rate (spectra/sec) 3.00
Instrument Parameters
Gas Temp
250 °C
Gas Flow
11 L/min
Nebulizer
40 psig
SheathGasTemp
350 °C
SheathGasFlow
12 L/min
Scan Segment
1
Scan Source Parameters
Vcap
3500 V
Nozzle Voltage
300 V
Fragmentor
120
Skimmer1
45
OctopoleRFPeak
750
ReferenceMasses
Ref Mass Enabled
Enabled
Use Bottle A
RefNebulizer
True
Ref Nebulizer
0 psig
To avoid problems in the LC/MS due to the high flow rate the effluent from the second dimension
column should be split. The recommended split ratio is 1:2
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...