![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575073.webp)
2D-LC User Guide
73
4
Installation
Hardware Installation
Figure 31
Stack Setup #1. Recommended setup if both pumps are Infinity II modules or the
2
D
pump is a 1290 Infinity Binary pump.
Number of
Capillary
#
Connection
L x ID [mm] P/N
Description
1a
1
1
D pump (top) to autosampler
400 x 0.17
5500-1245
Capillary ST 0.17x400 SI/SI
2
1
Autosampler to
1
D column (in MCT)
600 x 0.12
5067-4669
Capillary ST 0.12x600 S/SL
3a
1
1
D column to
1
D DAD
400 x 0.12
5500-1251
Capillary ST 0.12x400 SL/SL
4
1
1
D DAD to T-piece of PRK
400 x 0.17
5500-1245
Capillary ST 0.17x400 SI/SI
5
1
T-piece of PRK to Standard 2D-LC Valve
(Port 3) / ASM Valve (Port 4)
105 x 0.17
5500-1240
Capillary ST 0.17x105 SL/SL
6a
4
2D-LC Valve (1) - Deck (IN) – Deck (Out) -
2D-LC Valve (8) 2D-LC Valve (5) - Deck (IN) –
Deck (Out) - 2D-LC Valve (4)
170 x 0.12
5500-1270
Capillary ST 0.12x170 S/M
6b
4
ASM Valve (7) - Deck (IN) – Deck (Out) - ASM
Valve (3) ASM Valve (2) - Deck (IN) – Deck
(Out) - ASM Valve (8)
170 x 0.12
5500-1376
Capillary ST 0.12x170 M/M
(delivered with 2D-LC Valve
Kit, ASM)
7
1
2D-LC valve (6) / ASM valve (10) to
2
D
column (in MCT)
400 x 0.12
5500-1251
Capillary ST 0.12x400 SL/SL
8
1
2
D column (in MCT) to
2
D DAD
280 x 0.12
5067-4651
Capillary ST 0.12x280 SL/SX
9
1
2
D pump to 2D-LC Valve (7) / ASM Valve (5)
280 x 0.17
5067-4608
Capillary ST 0.17x280 SX/S
10
1
T-piece of PRK to damper capillary
150 x 0.17
5500-1227
Capillary ST 0.17x150 SL/SL
11
1
waste line
self-cut x
0.7
0890-1713
Tubing-flexible 0.8/1.61mm
PTFE WT (delivered with UV
detector)
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...