![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 182](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575182.webp)
6
Method Parameters
Additional Information
2D-LC User Guide
182
Example of a Dynamic Peak Parking Setup
Figure 98
Example picture of the dynamic cut shift setup
The maximum shift of the IRTS is 0.57 min. This means the end of peak-based
area is at 2.80 min and the next time-based cut with reference to IRTS can be
placed at 2.8 + 0.57 = 3.37 min.
Here is an example of how fluctuations that occur during a run can be
compensated with the help of the IRTS and dynamic cut shifting.
The chromatograms below shows the results obtained from MassHunter Qual.
The top 1D chromatogram shows the original reference chromatogram the
method was based on. The middle 1D chromatogram indicates the RT shift.
The cut-markers image below show that time-based cuts were dynamically
shifted accordingly.
Table 11
IRTS
Peak-based start:
2.10 min
Peak-based end:
2.80 min
Expected time:
2.67 min
NOTE
If a time-based cut shifts to the front and would enter the peak-based area
(bracket) the dynamic cut shift will not work.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...