2D-LC User Guide
369
14
Theoretical Background
Theoretical basis of 2D-LC
General
Definition
Alternatively peak capacity may be defined as the ratio of the total area A of the
chromatogram to the area A
0
required for the resolution of any zone:
n
c
defined that way is related to the geometrical definition by a factor:
Limits of Peak
Capacity in
2D-LC
Under ideal circumstances (
orthogonality
), the overall peak capacity (n
c,2D
)
should be equal to the product of the individual peak capacities of the first and
second dimension separations (
1
n
c
and
2
n
c
)
In practice the increase in peak capacity is not directly proportional to increase in
ability to resolve peaks.
Probable reason for this:
•
In 1D-LC, with a baseline width of a single component peak x
0
= 6
, x
0
units of
component free space on both sides of the maxima is necessary to ensure
baseline resolved peaks.
•
In 2D-LC the single component zone is A
0
= 2
r
2
and an area of component
free space of
(2r)
2
.
•
As a result: For every two component free widths in one dimension, four
component free areas are required in two dimensions.
Conclusions
for 2D-LC
1D-LC is inadequate for the separation of complex mixtures, as the number of
observable peaks compared to number of peaks to observe is too low. One
theoretical model (Statistical Model of Overlap = SMO), that correlates well with
real world observations, predicts, that the maximal fraction of the total peak
capacity that can be seen as chromatographic peaks is 37 % and even only 18 %
as single peaks. This implicates that extremely high peak capacities are needed
to separate complex samples with lots of components which is extremely
difficult to achieve.
Compared to 1D-LC separations, it's complicated to predict the number of
observable peaks in 2D-LC. For example, at a given peak capacity and a given
number of components, the aspect ratio in the two axes of separation has
impact on how effective the two separation are.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...