2D-LC User Guide
323
12
Maintenance
Correcting Leaks
Correcting Leaks
Correcting Leaks (G7116B)
1
Remove the door.
2
Use a pipette and tissue to dry the leak sensor area.
3
Observe the capillary connections and the column switching valve for leaks
and correct, if required.
4
Reinstall the door.
Correcting Leaks (G1170A)
1
Use a pipette and tissue to dry the leak sensor area.
2
Observe the capillary connections and the valve for leaks and correct, if
required.
When
If a leakage has occurred at the heat exchanger or at the capillary connections or at the column
switching valve.
Tools required
Description
Tissue
Pipette
Wrench, 1/4 – 5/16 inch
(for capillary connections)
When
If leakage has occured at the capillary connections or at the valve.
Tools required
Description
Tissue
Pipette
Wrench, 1/4 – 5/16 inch
(for capillary connections)