Information about the 5020TF fiber modules
Information about the 5020TF fiber modules
The 5020 TF single-mode and multi-mode contains two reels of fiber to be used
with the reflectometer. They are destined for training, demonstrations and can be
used as a launch fiber.
Note
As these modules are not qualified as measurement devices, there is
no need for calibration certificate.
Front panel of a module
Advice on how to use the fiber 2 (long fiber)
Given that the loss per unit on fiber 2 may be higher near port 1:
•
when measuring a dead zone, it is preferable to connect the reflectometer to
the port 4.
•
when measuring the insertion loss of a connector located after the launching
cable, the reflectometer should be connected to port 1.
Module 5020TF/MM
(Multi-mode)
Module 5020TF/SM
(Single-mode)
Type of fiber used
Corning MM
Corning SMF 28
Recommended group
index
1.50140 850nm ;
1.49660 @ 1310nm
1.46750 @ 1310nm ;
1.46810 @ 1550/1625nm
Length of fiber 1
Length of fiber 2
300
± 30 m
300
± 30 m
466
± 23 m
1000
± 50 m
Loss of the connectors
0.5 dB
a
a. Guaranteed to 1625 nm, except for the port 1 of fiber 2.
Port 1
Port 2
Port 3
Port 4