Remote control
MTS
e
: automatic printing of the results on the printer connected
to the MTS
e
at the time interval selected in seconds in the second
parameter.
REALtime
Choice of the real time or averaging mode. When this mode has
not be selected the power meter is, by default, in averaging mode.
ALArm
Selection of the required alarm threshold (-70 to +30 dBm).
ATTen
Entry of the value (0 to 30 db) of an external attenuator, if one is
being used.
REFeren
Select the currently displayed value as a reference value to
compare to other measured results.
ZERo
Adjust the zero setting of the power meter when a blanking cap is
fitted to the power meter optical input.
Power meter reference query
Format des commandes
Response: reference value for the Wavelength selected.
Laser source set-up
Format des commandes
Description des fonctions
LASer
ON: validates the emission from the laser or LED.
OFF: stops the emission from the laser or LED
LAMBda
Choice of wavelength (according to configuration of module):
1310 or 1550 nm for LASER source
850 or 1300 nm for LED source
PWMeter
GETRef?
[L850, L1310, L1550]
SOURce
LAMBda
MODe
POWer
[CW, M270, M330, M1000, M2000,SCANNING]
[CALIBRATED, VARIABLE], [0:255] , [0:255]
LASer
[ON, OFF]
[L850, L1300, L1310, L1550]
SENDlamb
EMUL
[NO, YES]