![background image](http://html1.mh-extra.com/html/acterna/mts-5100e/mts-5100e_user-manual_2842380193.webp)
U
SER
M
ANUAL
750000992/16
10-3
MTS
e
Battery replacement
Cleaning the optical cable connector
Note
If your module has an universal connector, unscrew its adaptor to
access the ferule.
•
Use a non degradable type of paper, such as Joseph paper, soaked in ethanol.
•
Pay particular attention to the polished face of the fiber, rubbing it in a
perpendicular motion to the fiber.
Cleaning the optical inputs or outputs of the MTS
e
•
Squirt a highly flammable solvent (such as ethanol or «Trijelt») into the
connector.
•
Blow out the connector using a clean dry air supply from an aerosol can fitted
with an extension.
Battery replacement
The MTS
e
lithium battery found on the microprocessor board ensures that date,
hour and certain information are not lost when the instrument is powered down.
Checking the battery
If there is doubt as to whether the battery is to be changed or not, switch off the
instrument. On switching the instrument on again, check whether the date is still
correct. If the date has been lost, the battery needs replacing.
Contact your nearest Wavetek Customer Service Centre.
This battery has a lifetime of more than five years.
Battery replacement should not be carried out by the user:
this may be dangerous to the microprocessor board
(certain components may be damaged or destroyed)
and requires special tools