![background image](http://html1.mh-extra.com/html/acterna/mts-5100e/mts-5100e_user-manual_2842380128.webp)
7-4
U
SER
M
ANUAL
750000992/16
Optical spectrum analyser
MTS
e
Results display parameters
Start screen
1
Start of display.
Auto
Automatic detection of start of display depending on the channels
detected. Start and end of displays are set automatically.
Full
Start of display at 1525 nm (for a 5071 WDM module).
Manual
Start of display manually chosen.
End screen
1
Auto
Automatic detection of end of display.
Full
End of display at 1570 nm (for a 5071 WDM module).
Manual
End of display manually chosen.
Tilt and Slope gain
No / Yes
Validates measurement and display of max. difference of gain (in
dB) and slope of gain (in dB by nm or by THz) above the trace.
Remarks in table
No / Yes
Allows display and entry of notes in table ("Table remarks" on
page 7-11).
Grid
Four different types of grid are available in the Display window of the menu:
Off
No
grid.
Convent
.
The grid coincides with the graduations. The interval therefore
varies according to the zoom level.
ITU
Press the <Set grid> key to access the ITU grid choices. Two keys
allow to have the ITU grid either only between the A and B cursors
or on the whole band.
Three lines in the menu allows the followings choices:
- ITU reference channel,
- interval (space) between channels: 50, 100 or 200 GHz.
- number of channels: 1 to 100.
Manual
Press the <Set grid> key to access a choice of the following grid
elements:
- first channel
- interval between channels:
1. The channels which are out of the screen limits are not in the result table.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com