Common part commands
•
For queries requiring parameters the “?” is between the last node and the first
parameter. They are given in the following command dictionary. A space is
required just after the “?”.
Examples:
FIB:K? L1310
request the K coefficient for the wavelength
1310 nm
ONOT? 24
request the note entered in the table for event
number 24
Common part commands
Remote Local mode (RS232 commands only)
Request to go to the remote mode.
Request to return to local mode.
Identification and option queries
MTS
e
request the MTS
e
identification.
MTS
e
Response: Acterna, MTS5100, n, X.YZ
with: n
serial
number
X.YZ software
version
Request options available on the MTS
e
Reset commands
Reset the SLGX1 LP to a predefined configuration (default state)
SRQ commands (IEEE interface only)
The IEEE.2 commands recognized by the MTS
e
are the following:
Device status register read request
Device Response
: NR1 number from 0 to 255.
Response to serial poll with bit 6: MSS.
Device service request enable
*SRE is followed by a number from 0 to 63 or 128 to 191.
Action
: The device modifies the bits in its service request enabling
register. This instruction is associated with *SRE?.
Device event register contents query
Device Response
: NR1 number from 0 to 255.
All events are cleared and the register is reset.
*REM
*LOC
*IDN?
*OPT?
*RST
*STB?
*SRE
*ESR?