5-26
U
SER
M
ANUAL
750000992/16
Using an OTDR module
MTS
e
Five cursor splice measurement result
3.
Successively press the <Expert>, then <Manual>, then <Splice> keys.
The splice event marker is positioned at the point defined by the cursor and
the result is displayed on the trace and in the event result table.
Important
To perform a five cursor splice measurement, when the <Splice> key
is pressed, either the A or B cursor must be selected, NOT the cursor
A/B.
Note
If no result is displayed this is due to: the display splice measure-
ment result threshold being higher than the splice you are trying to
measure, the threshold is set to
NO
splice display, or the Results on
trace function being disabled.
Note
If you try to measure a splice on a slope, the measurement is not car-
ried out and the “slope found in region between two cursors” error
message is displayed.
Manual ORL measurement
The ORL can be measured for a section of the fiber.
Use the following procedure to measure a section of the trace:
1.
Position the A and B cursors to define the section you want to measure.
2.
Successively press the <Expert>, then <Manual>, then <ORL> keys.