6-4
U
SER
M
ANUAL
750000992/16
Using the OTS module
MTS
e
instrument at the time interval selected with the Auto Time
function described below.
User Select
Selection of the any wavelength value between 800 nm to
1650 nm in 1 nm steps, using the arrow key <- or ->.
Time Auto
Choice of the time interval between measurements for
Autostore or Autoprint. This interval can be a period
(between 1 second and 11 hours 59 minutes and 59
seconds), in 1 second increments. It is selected by the
Arrow key <- and ->.
Real Time
Choice of the real time or averaging mode. When this mode
has not be selected the power meter is, by default, in
averaging mode.
Unit
Selection of the measurement unit:
W, dBm
or
dBr
.
Ext. Attenuator
Entry of the value (0 to 30 db) of an external attenuator, if
one is being used. It is selected by the Arrow key <- and ->.
Alarm Threshold
Selection of the required alarm threshold (-70 to +30 dBm).
OTS Results Window
The OTS Results window contains the Light source display and Power meter
display. These are only displayed if they have been selected from the Instrument
Setup window and if they are available.
Depending on the selection made they are either both shown together or only the
single selected unit.
The <Source> and <Power meter> softkeys select either the Light source or the
Power meter functions in the OTS results window.
A key also provides the results in memory to be displayed.
Note
When this memory is recalled with both the Power meter and Light
source windows, the Light Source window is reduced to a single line
at the bottom of the screen.