Laser source set-up
MODe
Transmission mode of signal emitted:
CW: continuous power
M270, M330, M1000, M2000: power modulated to selected
frequency (270 Hz, 330 Hz, 1 kHz, 2 kHz)
SCANNING: sources are alternated at regular intervals.
POWer
Transmitting power of laser source:
CALIBRATED
: power calibrated to 1 mW (0 dBm).
VARIABLE
: variable power from 0 to -10 dBm (if continuous mode
only).
Second parameter: rough setting from 0 to 255.
Third parameter: fine setting from 0 to 255.
SENDlamb
If continuous mode was chosen, this commands the sending of a
signal to the radiometer which switches it to the same wavelength
as the source.
EMUL
Used to emulate the operation of the OLS sources of the WWG
family.
Reading of OTS results
Format des commandes
Reply to command
PWE?
Power emitted by source in W or dBm.
MEA?
Power measured by radiometer in W, dBm or dBr.
MODe?
Reception mode: continuous (CW) or modulated to following
frequencies: 270 Hz, 330 Hz, 1 kHz or 2 kHz (M270, M330, M1000,
M2000).
BUFFer?
Result of measurement stored by Auto store mode. Response on a
line for each measurement:
Date Hour, Lambda, Result (dBm), result (dBr),
Received mode
Example of response:
22/12/98 09:42:28, 1310 nm, -10,03 dBm, 0,00 dBr, Cw
Only the parts of the buffer containing measurement results are
OTSResult
MEA?
MODe?
BUFFer?
PWE?
[W, DBM]
[W, DBM, DBR]
MEM?
[1:300]