What to do in the event of malfunctions
OPMI Lumera i
Version 8.0
Page 134
G-30-1720-en
Red reflex too dark or non-
existent.
Red reflex illumination not
activated.
Activate red reflex
illumination.
Contaminants on objective
lenses.
Clean objective lenses.
-
Light guide damaged or
incorrectly connected.
Check that light guide is
correctly seated and replace,
if necessary. Contact your
ZEISS Service.
Lamp is not properly seated in
the lamp socket.
Properly push lamp into lamp
socket.
Filter selection knob is in an
intermediate position.
Turn knob until the filter is
correctly in place.
Main microscope is not in the
correct working position.
Reposition the main
microscope.
Main microscope axis should
coincide with optical axis of
patient’s eye.
Image on video monitor too
dark or high level of noise.
Use of a camera adapter.
Remove fixed stop from
camera adapter.
Instructions
for Use for
camera
Activate the integral light
metering function of the
camera to reduce reflections
from the sclera and cornea.
Instructions
for Use for
camera
Increase the brightness level
of illumination.
Motor-driven focusing and/or
zoom functions of surgical
microscope are inoperative.
Stand electronics failure.
Manually adjust focusing
and/or zoom on surgical
microscope.
Motion of suspension arm
too stiff.
Friction adjustment knob
tightened too firmly.
Slightly loosen friction
adjustment knob.
No video image.
Connection cable not
properly connected.
Check all connections.
-
The stand is not stable.
Floor not level. Stand base
not appropriately positioned.
Slightly change the stand
base position.
-
Fault
Possible reason
Corrective measure
Cross
reference
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...