What to do in the event of malfunctions
OPMI Lumera i
Version 8.0
Page 136
G-30-1720-en
Zoom error. The error can be
acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., jammed hardware
If the error impairs your work,
switch to the manual mode.
Focus error. The error can be
acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., jammed motor of the
focus system
If the error impairs your work,
switch to the manual mode.
Error during system start.
Please reboot the system,
making sure that no function
buttons are pressed during
the booting process.
–
e.g., a button was pressed
during system start
Reboot the system. Make
sure not to press any buttons
during this process.
-
System error. The error can be
acknowledged.
If the error recurs, please
contact ZEISS Service.
–
Unknown software or
hardware error.
Acknowledge the error and
inform your ZEISS service
partner if it occurs again.
-
Error message
Possible reason
Corrective measure
Cross
reference
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...