What to do in the event of malfunctions
Malfunctions of the FCP foot control panel
Error messages on the 5.7" control panel
In the event of malfunctions occurring during operation, an error message is
displayed on the 5.7" control panel.
•
Check and eliminate the malfunction.
•
Acknowledge the error message.
The following table contains examples of error messages and the corresponding
remedies. If you have problems eliminating the malfunction or if errors keep
recurring, do not continue using the system and contact ZEISS Service.
You will find the contact responsible for your country on the following website:
www.meditec.zeiss.com
Fault
Possible reason
Corrective measure
Cross
reference
Foot control panel is
inoperative.
Connections on suspension
system or foot control panel
not plugged in.
Plug in connectors.
Error message
Possible reason
Corrective measure
Cross
reference
XY coupling error. The error
can be acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., hardware error in
motor electronics
–
e.g., software error,
inadmissible status of
motor control
If the error impairs your work,
switch to the manual mode.
Light source error. The error
can be acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., malfunction of light
intensity setting
If the error impairs your work,
switch to the manual mode.
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...