Device data
OPMI Lumera i
Version 8.0
Page 146
G-30-1720-en
Technical data
Essential performance features OPMI Lumera i
The OPMI Lumera i does not have any essential performance features as defined
in IEC 60601-1.
In lieu of testing the device’s essential performance features, certain device
functions defined by the manufacturer were tested.
The following device functions have been defined for the immunity test:
–
Visualization of the object must be maintained (minor image distortions
are permitted).
–
The lighting for object illumination must be continuously maintained
(interruptions of less than one minute are permitted).
–
The zoom must be adjustable.
–
Focusing must be possible.
–
Operation via the foot control panel must be maintained.
–
Operation via the control panel must be maintained.
–
The video image must be available.
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...