Safety measures
OPMI Lumera i
Version 8.0
Page 24
G-30-1720-en
Switch for manual mode
3
Switch for manual mode
In the event of failure of one of the major functions (XY movement, focus,
zoom, light control) and impairment of further functions, you can press
this switch to change to manual mode.
The system reacts as follows:
–
All electrical components except the light source are disconnected from
the power supply.
–
The light source adjusts to a medium brightness level to permit completion
of the application.
–
The 5.7" control panel is black.
To be able to continue surgery:
•
Manually set the magnification on the microscope’s zoom adjustment
knob (3.1) (e.g., with the help of a coin).
•
Manually position the suspension system for focusing and for moving
the system in the XY direction.
•
Use an external light source if the system light source set to a medium
brightness level is not sufficient.
Press switch (3) again to return to normal operation. The switch unlocks
and the system is restarted.
To ensure proper functioning of manual mode when required, check its
function at regular intervals (at least every six months) by activating the
function (without a patient).
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...