OPMI Lumera i
Device overview
Controls on the lamp housing
The lamp housing is designed for fiber illumination and is equipped with
a halogen light source. The lamp housing contains a backup lamp which
automatically swings into the illumination beam path when the first lamp fails.
1
Lamp module
2
Flap
for indicating the operating status of the halogen lamps
–
When the flap is closed, the main lamp is operative.
–
When the flap is open (yellow mark on the side), the backup lamp is
in use.
3
Manual switching to the backup lamp
4
Button for opening the lamp module
5
Filter selection knob
Fig. 13: Controls on
the lamp housing
No filter
Retina protection filter (blue barrier filter): when operating on
the eye, always use this filter to protect the patient’s eye against
unnecessary (blue) radiation.
KK 40 filter:
to increase color temperature
Empty filter position
1
2
3
4
5
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...