Device overview
OPMI Lumera i
Version 8.0
Page 38
G-30-1720-en
Controls on the microscope
1
DOF button (DeepView depth of field management system)
for selecting between optimum light transmission and maximum depth
of field. When deactivated (LED not lit), the microscope is optimized for
light transmission. When this function is activated (green LED is lit),
the microscope is automatically set to the optimum depth of field in
accordance with the selected magnification. The next time the system
is switched on, the mode last selected will be activated.
2
Zoom adjustment knob for manual mode
for manual controlling the zoom system.
3
Illumination knob
for selecting the illumination mode:
–
Position 1 – Red reflex illumination
This is the best setting to generate an optimum red reflex. Glare from the
sclera is greatly reduced as only the central field of view is illuminated.
–
Position 2 – Red reflex with surrounding field illumination
This setting permits clear visualization of the red reflex combined with
illumination of the surrounding field of view.
–
Position 3 – Surrounding field illumination
This setting is used for illuminating the field of view if no red reflex is
required.
–
Position 4 – Surrounding field illumination with retinal protection device
In this setting, a retinal protection device is swung into the surrounding
field illumination beam path. It prevents light from entering into the pupil,
providing additional protection for the patient’s eye against phototoxic
damage.
4
Light guide connector
Always make sure to insert the correct end of the light guide into the light
guide connector. For correct mounting, please see the label provided
below the light guide connector.
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...