Safety measures
OPMI Lumera i
Version 8.0
Page 20
G-30-1720-en
–
totally immobile eye, which means that only one area is subject to exposure,
–
pupil dilated to 8 mm.
The calculations are based on the daily exposure limits recommended for
occupational safety as defined in
6)
. A safety factor of 10 has been used in
determining these limits.
The use of the blue barrier filter increases the recommended exposure time by
a factor of 3 compared with the values specified below.
Maximum radiation exposure times with red reflex illumination
Maximum radiation exposure times with surrounding field illumination
Illumination display
scale in %
Max. radiation time in minutes
25
50
50
21
75
13
100
9
Illumination display
scale in %
Max. radiation time in minutes
25
8.7
50
3.7
75
2.2
100
1.6
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...