Failure of the halogen light source
The lamp housing contains a backup lamp which automatically swings into the
illumination beam path when the first lamp fails. The open flap (2) indicates that
the backup lamp is in operation. If the automatic swivel-in function is defective,
the backup lamp can also be swiveled in manually.
•
Press button (3) to manually swivel the backup lamp into the beam path.
•
Exchange the defective lamp at the first opportunity (see Page 140).
•
If the backup lamp is defective and replacement is not possible, continue
surgery using an external OR illuminator.
Fig. 36: Swiveling in the
backup lamp
NOTE
Overheating of the lamp module!
If ventilation grids (1) are covered, e.g., by drapes, this may lead to
overheating of the lamp modules and deactivation of the lamp.
•
Do not cover the ventilation grids.
1
2
3
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...